ACQUITY 通用平台样品管理器上出现 Z 和 Zp 轴错误(故障排除指导)- WKB98387

故障描述

  • 错误:Zp Axis Home H/W Fault(Zp 轴复位硬件出错)
  • 错误:Z Axis Home H/W Fault(Z 轴复位硬件出错)
  • 错误:Z Axis Move H/W Fault(Z 轴移动硬件出错)
  • 错误:Zp Axis Move H/W Fault(Zp 轴移动硬件出错)
  • 错误:Z axis calibration out of range(Z 轴校正值超出范围)
  • 错误:Zp axis calibration out of range(Zp 轴校正值超出范围)
  • 错误:Zp axis h/w lost steps(Zp 轴硬件丢失步骤)
  • 错误:Z axis h/w lost steps(Z 轴硬件丢失步骤)

环境

  • ACQUITY UPLC H-Class SM-FTN
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY Arc 样品管理器 FTN-R

策略

  1. 确保样品板配置设置正确。
    1. 对于 Empower - WKB56331
    2. 对于 UNIFI - WKB22166
  2. 寻找小车障碍物 - WKB2119
  3. 确保针没有损坏或弯曲 - WKB26140
  4. 确保针不会接触到样品瓶底部 - WKB11021

附加信息

 

id98387, A-30SM, axis error, needle carriage, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, 进样小车