NuGenesis Archive Agent 日志文件中出现错误“对注册表项 'HKEY LOCAL MACHINE\SOFTWARE\NuGenesis Technologies Corp.\NuGenesis\8.0\EVS\Vendor Registry\Waters Corp.' 的访问被拒绝” - WKB59006
故障描述
- 没有任何 Empower 项目归档到 NuGenesis
环境
- NuGenesis 8 SDMS
- NuGenesis 9 SDMS
- NuGenesis 8 Data Adapters
- Empower 3
原因
运行 NuGenesis RPC 服务的 Windows 帐户不是 Empower 远程计算机本地管理员组的成员。
解决方法
将该帐户添加到本地管理员组。
附加信息
Archive Agent(存档代理)的日志文件 Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log 中出现以下信息:
"E ","DATE TIME ","5 ","5 ","Access to the registry key 'HKEY_LOCAL_MACHINE\SOFTWARE\NuGenesis Technologies Corp.\NuGenesis\8.0\EVS\Vendor Registry\Waters Corp.' is denied. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","GetProjectList ","1 ",""
"E ","DATE TIME ","-1 ","-1 ","GetProjectList failed. ","Waters.SDMS.WindowsService ","clsEmp2PreProc ","PerformPreProcess ","1 ",""
"E ","DATE TIME ","-1 ","-1 ","error in Remote PreProcess method ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","DATE TIME ","-1 ","-1 ","PreProcessor call failed for the remote server: VCL ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""