NuGenesis 8 SDMS 未从 Empower 3 SR3 中存档 Empower 系统审计追踪 - WKB47289
故障描述
- NuGenesis 不备份 Empower 系统审计追踪
- 以下消息记录在 Archive agent(存档代理)的活动日志中:
"E ","datetime "(日期时间),"5 ","5 ","Argument 'Start' must be greater than zero. "(参数“开始”必须大于 0),"Waters.SDMS.WindowsService ","clsEmp2AuditPreProc ","GetProjectList ","1 ",""
"E ","datetime "(日期时间),"-1 ","-1 ","Backup maa files failed. "(maa 文件备份失败),"Waters.SDMS.WindowsService ","clsEmp2AuditPreProc ","PerformAuditPreProcess ","1 ",""
"E ","datetime "(日期时间),"-1 ","-1 ","error in Remote PreProcess method "(远程预处理方法错误),"Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
"E ","datetime "(日期时间),"-1 ","-1 ","PreProcessor call failed for the remote server: EMP3WIN10VM1 "(远程服务器预处理器调用失败:EMP3WIN10VM1),"Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
环境
- NuGenesis 8 SDMS
- Empower 3 Service Release 3
原因
Empower 3 SR3 中审计追踪的更改影响了 NuGenesis 中审计追踪备份过程的操作。
解决方法
截至 2018 年第 3 季度,尚无针对此问题的解决方法。
该缺陷已记录为 INFLMS-6145,并计划在 NuGenesis 9 中修复。
附加信息