e-SAT/IN WinSock 错误:(12) An instruction of Request Exclusive has failed((12)“请求独占”指令失败)- WKB222815
文章编号: 222815点击此处访问英文版本文章
故障描述
- 仪器出错 eSATIN#F18EST1182
- 仪器出错 eSATIN#F18EST1182:(5) The socket failed while waiting for incoming data with the following error code. WinSock Error #0((5) 等待传入数据时 socket 出错,并出现以下错误代码:WinSock Error #0)
- 仪器出错 eSATIN#F18EST1182:(6) A failure occurred in reading the reply from Get Data within is Data Available(在“数据是否可用?”中读取“获取数据”的回复时出错)
- 仪器出错 eSATIN#F18EST1182:(9) An attempt to open a communication path with the instrument has failed with error code. WinSock Error #10060((9) 尝试打开仪器的通讯路径失败,错误代码:Winsock error # 10060)
- 仪器出错 eSATIN#F18EST1182:(10) An instruction of Stop has failed to be sent((10) 发送停止指令失败)。
- 仪器出错 eSATIN#F18EST1182:(11) Sending information to the instrument failed with error code: WinSock Error # 10057((11) 发送信息到仪器失败,错误代码:WinSock Error # 10057)
- WinSock 错误:(12) An instruction of Request Exclusive has failed((12)“请求独占”指令失败)。
环境
- Empower 软件
- Windows 10
- e-SAT/IN 模块
- 直接以太网连接
原因
未确定,但在 Windows 7 上没有问题,只有在 Windows 10 上有问题。
解决方法
使用以太网交换机将 e-SAT/IN 连接到 LACE。
附加信息