ACQUITY UPLC 系统上出现鬼峰 - WKB200730

故障描述

  • 空白进样中出现鬼峰

环境

  • ACQUITY Arc
  • ACQUITY Arc Bio
  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS

原因

在线过滤器滤芯污染

解决方法

更换在线过滤器滤芯。

附加信息

订购信息:

  • 700002913,ACQUITY Arc
  • 700011679,ACQUITY Arc Bio
  • 700002913,ACQUITY UPLC
  • 700005173,ACQUITY UPLC H-Class
  • 700005259,ACQUITY UPLC H-Class Bio
  • 700002913,ACQUITY UPLC I-Class
  • 700005173,ACQUITY UPLC H-Class PLUS
  • 700005259,ACQUITY UPLC H-Class PLUS Bio
  • 700002913,ACQUITY UPLC I-Class PLUS