故障描述

  • 主动预加热器损坏,接口处渗漏,或管路破裂
  • 峰形异常,峰面积和峰高低于预期
  • 替换锥箍不适用于 APH

环境

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC 色谱柱管理器
  • ACQUITY UPC2

原因

频繁更换色谱柱导致 APH 管路破裂

解决方法

更换带有新装置的 APH。

  • H-Class APH = 205000730
  • H-Class Bio APH = 205001754
  • H-Class APH SS = 205001774
  • H-Class APH SS 18.5LG = 205000774
  • I-Class APH = 205000894
  • I-Class APH = 205001757
  • UPC2 APH = 205001002

附加信息

有关主动预加热器的完整列表,请参阅文章 WKB7232

 

id14176, A-30MCM, active preheater, UPC2CMAUX, UPCMA, UPCMAIVD, UPCMAUX, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPCMA, UPPCMAUX, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO