在 HDI 中发现 MaldiChrom 处理错误 - WKB46116

故障描述

  • 在 HD Imaging (HDI) 中处理数据时,观察到错误
  • 错误示例:

clipboard_eecaaca274d6e313fc9f4862b254cd918.png

MaldiChrom v2.14 Feb-2016(MaldiChrom v2.14 2016 年 2 月)
Copyright 2008-2016 Micromass UK Ltd.(版权所有 2008-2016 Micromass UK Ltd.)
[PROCPARS:gpps.cpp(445)] Failed to find key 'Tolerance' in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.(在文件 C:\HDImaging\process\MaldichromParameter.ini 的 [BACKSUB MALDICHROM PROCESS] 部分中无法找到“容差”项。)
[PROCPARS:gpps.cpp(445)] Failed to find key 'Flatten Edges’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.(在文件 C:\HDImaging\process\MaldichromParameter.ini 的 [BACKSUB MALDICHROM PROCESS] 部分中无法找到“使边缘平整”项。)
[PROCPARS:gpps.cpp(445)] Failed to find key 'Type’ in section [BACKSUB MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.(在文件 C:\HDImaging\process\MaldichromParameter.ini 的 [BACKSUB MALDICHROM PROCESS] 部分中无法找到“类型”项。)
[PROCPARS:gpps.cpp(435)] Failed to find section [TOFACM MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini. [PROCPARS:gpps.cpp(445]] Failed to find key ’Centroiding Type’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process \MaldichromParameter.ini.(在文件 C:\HDImaging\process\MaldichromParameter.ini 中无法找到 [TOFACM MALDICHROM PROCESS] 部分。在文件 C:\HDImaging\process \MaldichromParameter.ini 的 [CENTER MALDICHROM PROCESS] 部分中无法找到“质心类型”项。)
[PROCPARS:gpps.cpp(445]] Failed to find key ’Use Areas’ in section [CENTER MALDICHROM PROCESS] in file C:\HDImaging\process\MaldichromParameter.ini.(在文件 C:\HDImaging\process\MaldichromParameter.ini 的 [CENTER MALDICHROM PROCESS] 部分中无法找到“使用面积”项。)
0 lockmass adjustments to be applied(将应用 0 个锁定质量数调整)
MSE Level 0(MSE 级别为 0)

环境

  • HDI 软件
  • Synapt G2-S/Si MALDI

原因

这些消息是正常的,是由于未在 .ini 文件中指定参数而导致的。然后系统会使用缺省参数。

解决方法

无需执行任何操作。该消息正常。

附加信息

id46116, MALDIG2SI, MALDIG2SI, MALDIMICRO, SYNG2SIHD