WDM 的数据库备份日志中出现错误:Backup of the keystore failed(备份密钥库失败)- WKB252902
文章编号: 252902点击此处访问英文版本文章
故障描述
- WDM 的备份日志中出现以下信息:
- copy Drive:\ORACLE\PRODUCT\19.6.0\ADMIN\dbname\WALLET\tde\\ewallet.p12 Drive:\ORA_BKP\dbname\servicename\BACKUPSET\DATE_TIME\ewallet.p12 /Y (复制 Drive:\ORACLE\PRODUCT\19.6.0\ADMIN\dbname\WALLET\tde\\ewallet.p12 Drive:\ORA_BKP\dbname\servicename\BACKUPSET\DATE_TIME\ewallet.p12 /Y)
copy Drive:\ORACLE\PRODUCT\19.6.0\ADMIN\dbname\WALLET\tde\\cwallet.sso Drive:\ORA_BKP\dbname\servicename\BACKUPSET\DATE_TIME\cwallet.sso /Y(复制 Drive:\ORACLE\PRODUCT\19.6.0\ADMIN\dbname\WALLET\tde\\cwallet.sso Drive:\ORA_BKP\dbname\servicename\BACKUPSET\DATE_TIME\cwallet.sso /Y)
Backup of the keystore failed((备份密钥库失败))
- copy Drive:\ORACLE\PRODUCT\19.6.0\ADMIN\dbname\WALLET\tde\\ewallet.p12 Drive:\ORA_BKP\dbname\servicename\BACKUPSET\DATE_TIME\ewallet.p12 /Y (复制 Drive:\ORACLE\PRODUCT\19.6.0\ADMIN\dbname\WALLET\tde\\ewallet.p12 Drive:\ORA_BKP\dbname\servicename\BACKUPSET\DATE_TIME\ewallet.p12 /Y)
- 备份未完成
环境
- WDM 4.x
- Oracle 19c
- NuGenesis
- Empower 软件
原因
WATOracle 或 oraclejobs 用户不具备访问 Wallet 文件的权限。
解决方法
- 授予 WATOracle 或 oraclejobs(适用于 WDM 版本)对 ewallet.p12 和 cwallet.sso 文件的完全访问权限。
- 授予用户对备份集父文件夹的完全访问权限。
- 重新运行备份作业(如果使用 run_it_now 计划)或等待下一个计划的备份完成。
附加信息
id252902,