尝试在 NuGenesis LMS 中生成 PDF 报告时出错 - WKB98996

故障描述

  • 错误消息:An error occurred while receiving the HTTP response to http://servername/WATERSLMS/WebServi...ortService.svc. This could be due to the service endpoint binding not using the HTTP protocol. This could also be due to an HTTP request context being aborted by the server (possibly due to the service shutting down). See server logs for more details.(接收 http://servername/WATERSLMS/WebServi...ortService.svc HTTP 响应时发生错误。这可能是由于服务端点绑定未使用 HTTP 协议。这也可能是由于服务器中止了 HTTP 请求上下文(可能是由于服务关闭)。有关详细信息,请参阅服务器日志。)

环境

  • NuGenesis 8 ELN
  • NuGenesis 8 LMS
  • NuGenesis 9 LMS

原因

 WATERSELN PDF 打印机未在 ELN 应用程序服务器上正确加载。

附加信息

 

id98996, ELN, NGLMS, NGLMSLIC, NGLMSOPT, SUPNG