NuGenesis Archive Agent(存档代理)日志文件中出现错误:Error encountered while invoking remote EVS Server objects(调用远程 EVS 服务器对象时出现错误)- WKB58835
故障描述
- 没有任何 Empower 数据存档到 NuGenesis
环境
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- Empower 3 软件
原因
安装的 NuGenesis Data Adapters 与“远程 DCOM 服务器”计算机上安装的 Empower 3 版本不兼容。
解决方法
- 升级到适用于您所用 NuGenesis 版本的最新 Data Adapters 版本,因为与旧版本相比,最新版适配器可能支持更多版本的 Empower 3,并且其中包含许多错误修复和增强功能
- 如果最新 Data Adapter 版本不支持所安装的 Empower 版本,请考虑升级到更新版本的 Empower 3
附加信息
Archive Agent(存档代理)的日志文件“Drive:\Program Files (x86)\NuGenesis 8.0\AA\NGSDMS80AAActivity.log”中出现以下信息:
Message........: 2018-11-15 11:30:26: Beginning execution of archive request 1.
Message........: 2018-11-15 11:30:26: Server: SDMSQC, Project: OQFile_Emp_20181114, Template: Empower Archive 81.
Message........: 2018-11-15 11:30:26: Error encountered while invoking remote EVS Server objects. :Could not load file or assembly 'Oracle.DataAccess, Version=2.112.2.0, Culture=neutral, PublicKeyToken=89b483f429c47342' or one of its dependencies. The system cannot find the file specified..
Message........: 2018-11-15 11:30:26: Activitylog LogEntry: [NG14849] Failed to read field info list..
"E ","2018-11-15 11:30:26 ","5 ","5 ","Error encountered while invoking remote EVS Server objects. :Could not load file or assembly 'Oracle.DataAccess, Version=2.112.2.0, Culture=neutral, PublicKeyToken=89b483f429c47342' or one of its dependencies. The system cannot find the file specified.. ","Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
Message........: 2018-11-15 11:30:26: EMC: PerformPreProcess error.
Message........: 2018-11-15 11:30:26: Archive scan with no files archived for command id 1. {"SDMSQC","OQFile_Emp_20181114","Empower Archive 81"} {"Waters Corp.","Empower-3 Projects","0"}.{Summary : (Total Scanned Files = 0)(Start Time = 15-11-2018 11:30:26)(End Time = 15-11-2018 11:30:26)(Total Time = 0h 0m 0s)}.
Message........: 2018-11-15 11:30:26: Next scheduled runtime for archive request 1 is 2018-11-15 11:31:47.
id58835, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SDMS, SDMS8, SDMS8NU, SUP, SUPISDMS, SUPNG, upgrade, 项目