NuGenesis SDMS 未存档 Empower 系统审计追踪 - WKB57607

故障描述

  • NuGenesis 不备份 Empower 系统审计追踪
  • 以下消息记录在存档代理的活动日志中,具体取决于 SDMS 模板审计追踪是自动存档还是手动存档:
    • "E ","datetime "(日期时间),"5 ","5 ","Argument 'Start' must be greater than zero. "(参数“开始”必须大于 0),"Waters.SDMS.WindowsService ","clsEmp2AuditPreProc ","GetProjectList ","1 ",""
      "E ","datetime "(日期时间),"-1 ","-1 ","Backup maa files failed. "(maa 文件备份失败),"Waters.SDMS.WindowsService ","clsEmp2AuditPreProc ","PerformAuditPreProcess ","1 ",""
      "E ","datetime "(日期时间),"-1 ","-1 ","error in Remote PreProcess method "(远程预处理方法错误),"Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
      "E ","datetime "(日期时间),"-1 ","-1 ","PreProcessor call failed for the remote server: EMP3WIN10VM1 "(远程服务器预处理器调用失败:EMP3WIN10VM1),"Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""

    • "E ","datetime"(日期时间),"57 ","57 ","Logon failure: unknown user name or bad password.(登录失败:用户名未知或密码错误。)
       ","Waters.SDMS.WindowsService ","clsEmp2AuditPreProc ","GetProjectList ","1 ",""
      "E ","datetime"(日期时间),"-1 ","-1 ","Backup maa files failed. "(maa 文件备份失败),"Waters.SDMS.WindowsService ","clsEmp2AuditPreProc ","PerformAuditPreProcess ","1 ",""
      "E ","datetime"(日期时间),"-1 ","-1 ","error in Remote PreProcess method "(远程预处理方法错误),"Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""
      "E ","datetime"(日期时间),"-1 ","-1 ","PreProcessor call failed for the remote server: IBDCLIENT "(远程服务器预处理器调用失败:IBDCLIENT),"Waters.SDMS.EmpowerDAProxy ","clsMillPreProc ","PerformPreProcess ","1 ",""

  • 如果将 Empower 审计追踪手动备份到扫描路径,则 NuGenesis 可以存档审计追踪“.maa”文件

环境

  • NuGenesis 9 SDMS
  • NuGenesis 8 SDMS
  • Empower 3 软件

原因

  • 项目对象的 Empower 系统审计追踪策略需要注释和/或需要身份确认;
  • Empower 计算机操作系统中的“短日期”形式未设置为“mm/dd/yyyy”;
  • 由于计算机帐户的密码已过期,“Waters SDMS EmpowerDA DataExchange Service”无法通过域验证

解决方法

已针对此问题创建缺陷报告 CRI-207。

  • 在 Empower 中设置系统审计追踪策略:
    • 登录 Empower 配置管理器。
    • 单击“查看”>“系统策略”。
    • 选择“系统审计追踪”选项卡。
    • 为“项目”对象设置策略,如下所示:
      • “注释”:“安静”
      • “确认身份”=“未选中”
    • 单击“确定”
  • 在 Empower 客户端和服务器的操作系统中将短日期形式设置为 mm/dd/yyyy
  • 将计算机从域移到工作组中;重新启动计算机;将其添加回域。

附加信息

另请参阅:NuGenesis 8 SDMS 未从 Empower 3 SR3 中存档 Empower 系统审计追踪

id57607, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SDMS, SDMS8, SDMS8NU, SUP, SUPISDMS, SUPNG