NuGenesis SDMS Archive Agent(存档代理)错误信息:Rollback: Performing rollback for . . .(回滚:正在执行回滚 . . .)- WKB49103
故障描述
- NuGenesis 未捕获任何文件
- 在 Archive Agent(存档代理)日志文件中,出现以下错误信息:
Message........: DATE TIME: ExecuteMacroCommand: Could not restore the pending file (E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\70BCEC7E430FDD604B05978D9C591FB8). This file will not be archived.
Message........: DATE TIME: Rollback: Performing rollback for E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\5F5750854EC851FA7A31BA8A43A7678B.
Message........: DATE TIME: Rollback: Performing rollback for E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\54E70D504ABFC99846D607B17EC40000.
Message........: DATE TIME: Rollback: Performing rollback for E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\AD365FA946DEE923E707B99A24481319.
Message........: DATE TIME: Rollback: Performing rollback for E:\Program Files (x86)\NuGenesis 8.0\TE_TEMP\70BCEC7E430FDD604B05978D9C591FB8.
Message........: DATE TIME: ExecuteTemplateCommand: The group is rolled back because of errors.
环境
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
原因
系统硬盘中的所有可用空间均已用尽,无法再将 AA 捕获的组中文件复制到 TE_TEMP。
解决方法
无。File Capture(文件捕获)系统将尝试在后续运行中捕获仪器数据。
附加信息
如果 FC 服务器经常出现空间不足,可能需要增加 TE_TEMP 文件夹配置硬盘驱动器的大小。