NuGenesis SDMS 数据管理模块仅尝试删除包含后代项目的 Empower 项目一次,删除失败并显示 Deletion Failed with Error - Captured Data not integral(删除失败并出错 - 采集的数据不完整)- WKB247185
文章编号: 247185点击此处访问英文版本文章
故障描述
- 如果由于正在删除的 Empower 项目中有尚未删除的子项目而导致 Empower 项目删除失败,VISION 中的删除状态显示为 Deletion Failed with Error - Captured Data not integral(删除失败并出错 - 采集的数据不完整),之后不再尝试
- DM Delete(DM 删除)状态为 FINAL(最终),未再次尝试删除
- NGCONMTENTMASTER 表中对应的 NGDMDELETEFLAG 值为“6”
- 在 DMM 日志中,您会看到如下所示的错误:
[7] : 2022-11-03 14:19:11:Initializing...
[] : 2022-11-03 14:19:11:Inited Successfully : Waters Corp. Empower-Millennium Projects DM EVS Plug~n~Play object!!
[7] : 2022-11-03 14:19:11:Data Management scan for Project [Empower Projects], Vendor [Waters Corp.] Product [Empower-3 Projects] Revision [0] started at [ 03-Nov-22 14:19:11].
[7] : 2022-11-03 14:19:12:Duration of candidate scan: 0h 0m 1s.
[] : 2022-11-03 14:19:21:Not Deleted Project:[EMP3.DMM Order\B\Tea 01] Remote machine: [WB-EMP-CL] it has descendants - Err.=-2147220839 – handle the project manually
环境
- NuGenesis SDMS,所有版本
原因
这是一个软件缺陷,已提交为 CRI-5171。
解决方法
暂无解决方法。SDMS 数据管理删除后代项目后,需要手动删除 Empower 父项目。
附加信息
id247185, SDMS, SDMS8, SDMS8NU, SUPISDMS, SUPNG