在 Empower 中观察到错误:Agilent LC Setup has failed, due to discrepancies between method configuration and on-line instrument configuration(由于方法配置和在线仪器配置不符,Agilent LC 安装失败)- WKB12279
文章编号: 12279点击此处访问英文版本文章
故障描述
错误:Agilent LC Setup has failed, due to discrepancies between method configuration and on-line instrument configuration(由于方法配置和在线仪器配置不符,Agilent LC 安装失败)
错误:Agilent LC Communication: A Configuration Change Event has been reported. An instrument has been removed or added to the current system. It is recommended that you Re-Identify your system to detect the new configuration(Agilent LC 通讯:已报告“配置改变事件”。有仪器被移走或添加到当前系统。建议重新识别系统来检测新配置。)
环境
- Empower 2
- Empower 3
- Agilent LC 控制软件 v1.06.1007
- A1100 固件 vA.06.04
- 系统定义为 A1100
原因
PCS# 47944
修复/解决方法
在 Agilent Instrument Control Framework (ICF) 中,转到仪器控制,将系统定义为 AgilentLC,而不是 A1100。
- 要求 ICF 版本不低于 v2.13
附加信息
id12279, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, firmware, SUP, 安捷伦