在 Empower 中观察到错误:Agilent LC Setup has failed, due to discrepancies between method configuration and on-line instrument configuration(由于方法配置和在线仪器配置不符,Agilent LC 安装失败)- WKB12279

故障描述

  • 错误:Agilent LC Setup has failed, due to discrepancies between method configuration and on-line instrument configuration(由于方法配置和在线仪器配置不符,Agilent LC 安装失败)

  • 错误:Agilent LC Communication: A Configuration Change Event has been reported. An instrument has been removed or added to the current system. It is recommended that you Re-Identify your system to detect the new configuration(Agilent LC 通讯:已报告“配置改变事件”。有仪器被移走或添加到当前系统。建议重新识别系统来检测新配置。)

环境

  • Empower 2
  • Empower 3
  • Agilent LC 控制软件 v1.06.1007
  • A1100 固件 vA.06.04
  • 系统定义为 A1100

原因

PCS# 47944

修复/解决方法

在 Agilent Instrument Control Framework (ICF) 中,转到仪器控制,将系统定义为 AgilentLC,而不是 A1100。

  • 要求 ICF 版本不低于 v2.13